Second Generation Venous Probe for Cardio Help VP02CH

Second Generation Venous Probe for Cardio Help VP02CH
custom

Second Generation Venous Probe for Cardio Help VP02CH

for Maquet Cardiopulmonary GmbH

Application field : Cardiac SurgeryCardiopulmonary supportECMOEmergency medicine
Measured Parameters : Blood temperature (Tblood)Hematocrit (Hct)Hemoglobin (Hb)Oxygen Saturation (SO2)
Sensors type : Handheld

Description

VP02CH is the venous probe used with the Cardiohelp System made by Maquet. 

The Cardiohelp System is a compact heart-lung support system suitable for all indications requiring extracorporeal circulation for cardiopulmonary support.

The VP02CH sensor together with the system is used for monitoring blood parameters through non-invasive optical methods.

The Cardiohelp system together with the VP02CH sensor can be used for up to 30 days in cases of myocardial or respiratory failure.

Range, Accuracy & Resolution
ParameterRangesAccuracyResolution
Hematocrit
15 – 45 % Hctbetter or equal to ± 5 Hct units
0.1%
Hemoglobin
5 – 15 g/dl± 1,5 g/dl 0.1 g/dl
Oxygen saturation40 – 99.9 % O2Sat
± 5 O2Sat units0.1%
Blood temperature range10 – 45 °C±0.7°C for blood temperature range +24 – +37°C and ambient temperature range +20 – +24°C;0.1°C
±1.0°C for blood temperature range +24 – +37°C and ambient temperature range +15 – +30°C
Outside the mentioned ranges larger measurement inaccuracies are possible.
Technical Details
Cuvette connector sizeMin flow rateMax flow rate
3/8” and 1/4”0.5 l/min7 l/min
Communication configuration
Serial SR232
Host equipment
  • Cardiohelp System of Maquet Cardiopulmonary GmbH (GETINGE Group)