OVP — Optical Venous Probe

Optical Venous Probe OVP, top view
Optical Venous Probe OVP, bottom view with tubing
Optical Venous Probe OVP, bottom view without tubing
custom

OVP — Optical Venous Probe

Custom design for LivaNova PLC

Application field : Cardiac SurgeryCardiopulmonary supportECMO
Measured Parameters : Blood temperature (Tblood)Hematocrit (Hct)Hemoglobin (Hb)Oxygen Saturation (SO2)
Sensors type : Handheld

Description

OVP is the Optical Venous Probe used with the B-Capta system made by LivaNova.

The OVP sensor together with the system provides an in-line continuous monitoring of the patient’s blood parameters. It is used in the B-Capta system which is designed to easily and accurately monitor Venous Blood Gas Parameters even in long and complex Pediatric and Adult cardiopulmonary bypass procedures.

OVP is the Optical Venous Probe used with the B-Capta system made by LivaNova.
The OVP sensor together with the system provides an in-line continuous monitoring of the patient’s blood parameters:

  • blood oxygen saturation,
  • hematocrit,
  • hemoglobin
  • temperature.
Range, Accuracy & Resolution
ParametersRangesAccuracyResolution
Hematocrit13 – 55 % Hct ± 3 Hct units0.1%
Hemoglobin4.1 – 18 g/dl± 1 g/dl 0.1 g/dl
Oxygen saturation40 – 99.9 % O2Sat± 5 O2Sat units± 0.1%
Blood temperature range15 – 41 °C± 1°C0.1°C
Technical Details
Cuvette connector sizeMin flow rateMax flow rate
1/2”0.5 l/min
8.0 l/min
3/8”0.2 l/min 6.0 l/min
1/4”0.1 l/min 2.5 l/min
Communication configuration
RS232
Host equipment
  • B-Capta of LivaNova PLC