available-technology

Sweep Gas Carbon Dioxide Sensor

available-technology

Sweep Gas Carbon Dioxide Sensor

Application field : Cardiac SurgeryCardiopulmonary supportECMOEmergency medicineOrgan detoxificationResuscitationTherapies of CO2 removal
Measured Parameters : Carbon Dioxide (PECO2)
Sensors type : Handheld

Hightlight

Disposable Adapter for Maximum Safety – designed with a dedicated disposable adapter for the exhaust gas line of the oxygenator. This solution eliminates the need to sterilize the sensor probe between uses, eliminating cross-contamination risks and streamlining clinical workflows

Hot and Humid Gas Tolerance – the sensor is exceptionally resilient to hot, moisture-saturated gas flows. It guarantees stable, continuous performance without signal degradation or interference from condensation.

High-Concentration Measurement Capability – built for advanced biomedical applications, the device accurately measures high CO2 levels well above indoor air quality ranges, covering up to 4% – 5% volume (40,000 – 50,000 ppm).

Digital Interface for Real-Time VCO2 – features a digital I2C interface that provides a direct signal for seamless microcontroller integration. This enables the continuous, real-time calculation of CO2 removal (VCO2) by combining the measured gas fraction with the sweep gas flow rate.

Integrated Climate Monitoring – To ensure absolute precision and environmental data tracking within the gas line, the module includes a fully integrated relative humidity and temperature (RH/T) sensor.

Description

The Sweep Gas Carbon dioxide Sensor is designed to be integrated into the exhaust gas line of a membrane lung.

The sensor measures the CO2 fraction in the sweep gas exhaust, which, combined with the sweep gas flow rate, enables the continuous calculation of CO2 removal (VCO2).

Technical Data
Parameter DimensionsData
CO2Measurement Range0 –50,000 ppm (0-5% vol.)
Power supply2.4 V –5.5 V
Operating Humidity0 –95% RH (Non-condensing)
Accuracy±(50 ppm + 5% of reading)
Operating Temperature-10°C to +60°C
Digital InterfaceI2C
RH/T Sensorintegrated