O2PP — Optical Sensor for Oxygen Partial Pressure

available-technology

O2PP — Optical Sensor for Oxygen Partial Pressure

Application field : HemodialysisPlasmapheresis
Sensors type : Panel Mounted

Unique features

Calibration free – Sensor is fully ex-vivo blood calibrated in house before shipment

Non-invasive continuous monitoring of oxygen partial pressure in blood

No calculation is required at the host equipment level, physiological values with different measurement units can simply be displayed

Possibility to align the reading values with a specific customer reference equipment

Description

The O2PP sensor probe is aimed for measuring the oxygen partial pressure of blood, by Neural Network in an extra-corporeal blood circuit, so it can be applied for monitoring blood parameters during therapies requiring extra-corporeal  blood circulation such as haemodialysis, plasmapheresis.

Range, Accuracy & Resolution
ParameterMeasured Ranges/Accuracy
Oxygen partial pressure pO2 range50–400 mmHg
Blood temperature range 15.0–40.0 °C
pO2 at actual °C accuracy±10% mmHg for 90–400 mmHg
±15% mmHg for 50–90 mmHg
pO2 at 37 °C accuracy±10% mmHg for 90–400 mmHg
±15% mmHg for 50–90 mmHg
Blood temperature accuracy±1.0 °C
Ambient temperature accuracy15 °C to +40 °C
Blood flow rate*According to cuvette size
3/8" flow rate [0.2 – 6.0] LPM
1/4" flow rate [0.1 – 2.5] LPM
* 3/8″ and 1/4″ cuvettes sizes available
Technical Details
ParameterDescription
Power supply+5V ±5% @100mA
+12V ±5% @80mA
Storage temperature disposableSensor spot must be stored at room temperature, away from heating sources and from direct sunlight, in the original black shielding pouch which protect the spot from the ambient light.
LifetimeProved for at least 24 hours
Shelf-life1 year
Response timeT90<60 s for all measured and calculated parameters
Communication interfaceRS232 - Baud Rate 9.600
Transport sterilisation conditionsThe sensor spot withstands a temperature range of -30°C/+55°C. Ethylene Oxide sterilization is the only validated method of choice
Communication protocolMaster-Slave type. The protocol is equipped with a 16-bit CRC for every transmitted frame.

Measurement Principle

The oxygen partial pressure is measured in real-time by acquisition of oxygen-quenched optical fluorescence decay signal*

* Patent pending PCTIB2022051680_WO2022200875A1

Fluorescent process in absence of Oxygen

Interrogation of a dye with UV light Fluorescent emission decay over time

Fluorescent process in presence of Oxygen

The rate of fluorescence decay velocity varies depending on the oxygen concentration

Oxygen Increasing concentration