Calibration free – Sensor is fully ex vivo blood calibrated in-house before shipment
Easily and directly coupled to the extracorporeal tubing line
Non-invasive continuous monitoring of blood Saturation, blood temperature, and haematocrit
No calculation is required at the host equipment level, physiological values can simply be displayed
Possibility to adjust the measured values with laboratory values
Description
This SatCrit Probe is aimed at measuring a plurality of blood parameters (Sat%, Hct%, Hb, and Blood Temperature) by Neural Network in an extra-corporeal blood circuit.
Application is for monitoring blood parameters during therapies requiring extra-corporeal blood circulation, such as haemodialysis, plasmapheresis, extra-corporeal respiratory assistance (ECMO), the preserva on of explanted organs, and cancer therapies.
Measurement Principle
The plurality of blood parameters is monitored in real time employing Neural Network*. Blood is non invasively “interrogated” with electromagnetic radiation at a plurality of determined wavelengths, and a plurality of light responses, being retro-reflected or diffused by blood, are detected.
*Patent pending PCT/IB2023/060787_WO2024089628A1
Range, Accuracy & Resolution
Parameter
Measured Range / Accuracy
Blood Flow Range
0,5 – 7 LPM according to 3/8" tubing size
Blood Temperature Range
15.0 – 40.0 °C
Blood temperature accuracy
± 1.0 °C
Haematocrit range
15,0 - 45,0 %
Haematocrit accuracy
±5.0 points
Haemoglobin range
5.0– 15.0 [g/dl]
Haemoglobin accuracy
±1.65 [g/dl]
Blood Saturation range
40.0 – 99.9 %
Blood Saturation accuracy
±5.0 points
Ambient temperature range
15°C to +40°C.
Technical Details
Parameter
Value
Power supply
+5V ±5% @250mA
Communication Interface
Serial RS232 - Baud rate 9.600
Response time
T90<60 s for all measured and calculated parameters