available-technology

SatCrit - Optical Sensor for Blood Oxygen Saturation & Haematocrit, direct on standard tube

available-technology

SatCrit - Optical Sensor for Blood Oxygen Saturation & Haematocrit, direct on standard tube

Application field : ECMOHemodialysisPlasmapheresis
Measured Parameters : Blood temperature (Tblood)Hematocrit (Hct)Hemoglobin (Hb)Oxygen Saturation (SO2)

Unique features

Calibration free – Sensor is fully ex vivo blood calibrated in-house before shipment

Easily and directly coupled to the extracorporeal tubing line

Non-invasive continuous monitoring of blood Saturation, blood temperature, and haematocrit

No calculation is required at the host equipment level, physiological values can simply be displayed

Possibility to adjust the measured values with laboratory values

Description

This SatCrit Probe is aimed at measuring a plurality of blood parameters (Sat%, Hct%, Hb, and Blood Temperature) by Neural Network in an extra-corporeal blood circuit.

Application is for monitoring blood parameters during therapies requiring extra-corporeal blood circulation, such as haemodialysis, plasmapheresis, extra-corporeal respiratory assistance (ECMO), the preserva on of explanted organs, and cancer therapies.

Measurement Principle

The plurality of blood parameters is monitored in real time employing Neural Network*. Blood is non invasively “interrogated” with electromagnetic radiation at a plurality of determined wavelengths, and a plurality of light responses, being retro-reflected or diffused by blood, are detected.

 

*Patent pending PCT/IB2023/060787_WO2024089628A1
Range, Accuracy & Resolution
ParameterMeasured Range / Accuracy
Blood Flow Range0,5 – 7 LPM according to 3/8" tubing size
Blood Temperature Range15.0 – 40.0 °C
Blood temperature accuracy± 1.0 °C
Haematocrit range15,0 - 45,0 %
Haematocrit accuracy±5.0 points
Haemoglobin range5.0– 15.0 [g/dl]
Haemoglobin accuracy±1.65 [g/dl]
Blood Saturation range40.0 – 99.9 %
Blood Saturation accuracy±5.0 points
Ambient temperature range15°C to +40°C.
Technical Details
ParameterValue
Power supply+5V ±5% @250mA
Communication InterfaceSerial RS232 - Baud rate 9.600
Response timeT90<60 s for all measured and calculated parameters
Blood flow rateAccording tube size
3/8 inch: flow rate [0.5 ÷ 7.0] LPM
Transport conditionsThe sensor withstands a temperature range of -25°C/+60°C
Communication protocolMaster-Slave type, equipped with a 16-bit CRC for every transmitted frame
Tubing size**Plasticized PCV - ID 3/8" wall thickness 3/32"
Checked Tubing compatibilityECC no DOP Bloodline 9,5x2,4 mm MED 7536 of Raumedic AG
Nalgene™ Non-Phthalate PVC Tubing (code 8701-0130) of Thermo Scientific
** On request design for other tubing sizes is possible