Calibration free – Sensor is fully ex-vivo blood calibrated in house before shipment
Non-invasive continuous monitoring of oxygen partial pressure in blood
No calculation is required at the host equipment level, physiological values with different measurement units can simply be displayed
Possibility to align the reading values with a specific customer reference equipment
The O2PP sensor probe is aimed for measuring the oxygen partial pressure of blood, by Neural Network in an extra-corporeal blood circuit, so it can be applied for monitoring blood parameters during therapies requiring extra-corporeal blood circulation such as haemodialysis, plasmapheresis.
| Parameter | Measured Ranges/Accuracy |
|---|---|
| Oxygen partial pressure pO2 range | 50–400 mmHg |
| Blood temperature range | 15.0–40.0 °C |
| pO2 at actual °C accuracy | ±10% mmHg for 90–400 mmHg ±15% mmHg for 50–90 mmHg |
| pO2 at 37 °C accuracy | ±10% mmHg for 90–400 mmHg ±15% mmHg for 50–90 mmHg |
| Blood temperature accuracy | ±1.0 °C |
| Ambient temperature accuracy | 15 °C to +40 °C |
| Blood flow rate* | According to cuvette size 3/8" flow rate [0.2 – 6.0] LPM 1/4" flow rate [0.1 – 2.5] LPM |
| Parameter | Description |
|---|---|
| Power supply | +5V ±5% @100mA +12V ±5% @80mA |
| Storage temperature disposable | Sensor spot must be stored at room temperature, away from heating sources and from direct sunlight, in the original black shielding pouch which protect the spot from the ambient light. |
| Lifetime | Proved for at least 24 hours |
| Shelf-life | 1 year |
| Response time | T90<60 s for all measured and calculated parameters |
| Communication interface | RS232 - Baud Rate 9.600 |
| Transport sterilisation conditions | The sensor spot withstands a temperature range of -30°C/+55°C. Ethylene Oxide sterilization is the only validated method of choice |
| Communication protocol | Master-Slave type. The protocol is equipped with a 16-bit CRC for every transmitted frame. |