BLD&DDM — Combined Blood Leakage Detector + Dialysis Dose Monitor Sensor

available-technology

BLD&DDM — Combined Blood Leakage Detector + Dialysis Dose Monitor Sensor

Application field : Blood leakage detectionHemodialysis
Sensors type : Panel Mounted

Unique features

100% factory calibration

Adoption of a robust quartz cuvette

Data output includes: Blood concentration, Blood leakage threshold determination, Monitor of UV and Near IR absorption, Monitor of Disinfection Fluid’s temp.

Description

The combined BLD & Dialysis Dose Monitor is a non invasive optical sensor probe designed to detect the smallest amounts of blood, as well as the concentration of urinary excreted substances in the spent dialysate fluid.

The sensor probe adopts two light sources in the Ultraviolet spectrum range and one light sources in the near Infrared spectrum range in order to provide the best detection sensitivity for both the blood leak (according to IEC 60601-2-16:2008) and Uric Acid concentration.

 

*Quartz Cuvette not provided
Technical Data
Technical SpecificationValue
Data communicationserial communication via RS232 or USB
Power supply+5V ±5% @200mA
Fluid temperature25 – 90 [°C]
Expected Service Life10 years
Storage temperature-20 – +60 [°C]
Response Time5 [s] at T90
StandardsSafety req.: IEC 60601-2-16:2008
EMC: IEC 60601-1-2:2014
Operating temperature10 – 50 [°C]
Dialysate Flow rate range0,1 – 2,0 [l/min]
Range
Parameter BLD & DDMValue
Change in UV absorption0.1–3 [arbitrary units]
which correlates to decrease in Uric Acid concentration
spKT/V and URRcalculated by the HD equipment from UV absorption data
Blood concentration in dialysate0,05 – 1 [‰]
Sensitivity0,05 [‰]
BLD Status Code0: no blood leak
1: blood leak
Dialysate Temperaturerange 25°C - 90°C, accuracy ± 2°C

Optical performance

Optical performance of BLD section

On-line curve where NIR/UV absorbance is plotted versus blood concentration in dialysate

Optical performance of DDM section

A typical on-line curve where UV-absorbance is plotted versus time.

The evolution of Kt/V parameter is estimated minutes by minute either during treatment and confirmed at the end of the treatment from the slope of the logarithmic plot of the on-line absorbance values.
Here on the graph the slope of the absorbance fitting curve Sa is -0.0055, so the Kt/V values for a 270 minutes long dialysis can be obtained as (Kt/V)a = – Sa *t = 0.0055*270 = 1,48