available-technology

DDM — Optical Dialysis Dose Monitor Sensor

available-technology

DDM — Optical Dialysis Dose Monitor Sensor

Application field : Hemodialysis

Unique features

100% factory calibration.

Adoption of a robust quartz cuvette.

Data output: Monitor of UV absorption.

Description

The Dialysis Dose Monitor is a non-invasive optical sensor probe designed to detect the concentration of urinary excreted substances in the spent dialysate fluid.
The sensor probe adopts a light source in the Ultraviolet spectrum range in order to provide the best detection sensitivity of Uric Acid concentration in the spent dialysate fluid.

 

*Quartz Cuvette not provided

 

Technical Details
Technical SpecificationValue
Data communicationserial communication via RS232
Power supply+5V ±5% @200mA
Fluid temperature25 – 90 [°C]
Expected Service Life10 years
Storage temperature-20 – +60 [°C]
Response Time!ERROR! B7 -> Table for Structured Reference cannot be identified
StandardsSafety req.: IEC 60601-2-16:2008
EMC: IEC 60601-1-2:2014
Operating temperature10 – 50 [°C]
Dialysate Flow rate range0,1 – 2,0 [l/min]

 

Range
Parameter DDMValue
Change in UV absorption0.1–3 [arbitrary units]
which correlates to decrease in Uric Acid concentration
spKT/V and URRcalculated by sensor from UV absorption data
Optical performance of DDM section

A typical on-line curve where UV-absorbance is plotted versus time.
The evolution of Kt/V parameter is estimated minutes by minute either during treatment.